ZS3 Instructions for Use
138
|
P a g e
NOTE:
1 The system needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided below.
2 Other devices may interfere with this system even though they meet the
requirements of
CISPR
.
3 Preventing conducted RF immunity. Due to technological limitations, the conducted
RF immunity level are limited to 3Vrms level, conducted RF interference above 3Vrms
may cause wrong diagnosis and measurements. We suggest that you position system
further from sources of conducted RF noise.
4 Portable and mobile RF communications equipment can affects system. See tables 1,
2, 3, and 4 below.
5 The EMISSIONS characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for
which CISPR 11 class B is normally required) this equipment might not offer adequate
protection to radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.
If the system is operated within the electromagnetic environment listed in Table 2 and Table 3, the
system will remain safe and will provide the following basic performances:
Imaging;
Doppler acoustic spectral displaying;
Taking measurements;
Patient information;
Date/time information.
TABLE 1
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC EMISSIONS
The system is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
EMISSIONS TEST
COMPLIANCE
ELECTROMAGNETIC ENVIROMENT
-
GUIDANCE
RF emissions
CISPR 11
Group 1
The system uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A
The system is suitable for use in all
establishments including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/ Flicker
Emissions
IEC 61000-3-3
Compliance
Содержание Zonare ZS3
Страница 1: ...I ZS3 Ultrasound System Instructions for Use ...
Страница 6: ......
Страница 12: ......
Страница 38: ......
Страница 56: ...ZS3 Instructions for Use 48 P a g e Figure 4 6 P8 3m Adult and Pediatric TEE Scan Plane adjustment ...
Страница 74: ......
Страница 78: ......
Страница 86: ......
Страница 110: ......
Страница 128: ......
Страница 146: ......
Страница 150: ......
Страница 158: ......
Страница 160: ......
Страница 164: ...ZS3 Instructions for Use 147 P a g e ...
Страница 165: ...P N 046 012347 01 3 0 ...