ZS3 Instructions for Use
52
|
P a g e
Wearing the Probe Sheath
A legally marketed probe sheath must be installed over the probe before performing intra-cavitary
and intra-operative examination. Protective barriers may be required to minimize disease
transmission. Probe sheaths are available for use with all clinical situations where infection is a
concern.
To order probe sheath, contact:
CIVCO Medical Instruments Co.
102 First Street South, Kalona, IA 52247-9589 USA Tel: 1-319-656-4447
E-mail: [email protected]
http://www.civco.com
CAUTION:
1. Be sure to cover the probe with a new (unused) probe sheath to prevent
infection during examination. If the package of a probe sheath is open or
broken, the sterilization of the probe sheath may not be sufficient. DO
NOT use such a probe sheath.
2. The cover contains natural rubber latex and talc that can cause allergic
reactions in some individuals.
3. DO NOT use an expired probe sheath. Before using a probe sheath, verify
whether the term of validity has expired.
4. For neurological intra-operative probes, users are recommended to use a
sterile, non-pyrogenic sheaths.
5. For neurological intra-operative probes, user should be aware of the
following potential problem in using the probe on patients with known or
suspected Creutzfeldt Jakob disease (CJD). The probe sheath should not
be relied upon to prevent contamination of the probe. A transducer
exposed to central nervous system tissue from known or suspected CJD
or variant CJD (vCJD) should be destroyed since it may not be possible to
sterilize it.
Method (for reference only):
1.
Place an appropriate amount of gel inside
the sheath or on the probe acoustic lens.
Poor imaging may result if no gel is used.
2.
Insert the probe into the sheath; make
sure to use proper sterile technique. Pull
cover tightly over probe acoustic lens to
remove wrinkles and air bubbles, and
taking care to avoid puncturing the
sheath.
Содержание Zonare ZS3
Страница 1: ...I ZS3 Ultrasound System Instructions for Use ...
Страница 6: ......
Страница 12: ......
Страница 38: ......
Страница 56: ...ZS3 Instructions for Use 48 P a g e Figure 4 6 P8 3m Adult and Pediatric TEE Scan Plane adjustment ...
Страница 74: ......
Страница 78: ......
Страница 86: ......
Страница 110: ......
Страница 128: ......
Страница 146: ......
Страница 150: ......
Страница 158: ......
Страница 160: ......
Страница 164: ...ZS3 Instructions for Use 147 P a g e ...
Страница 165: ...P N 046 012347 01 3 0 ...