ZS3 Instructions for Use
126
|
P a g e
possible.
•
Select Color Doppler mode.
•
Wait for 2 minutes.
•
Feel at the distal end of the probe if there is a temperature increase that could be harmful for
the patient.
•
When a brush is used for cleaning the transducer, use only a soft brush; coarse/stiff- bristle
brushes may cause transducer damage
•
Keep the control handle and transducer connector out of any cleaning or disinfection
solutions.
•
The control handle and cable may be cleaned with a damp cloth, but only the distal end of the
probe up to the 100cm marker on the shaft may be placed into a disinfection solution.
•
During immersion disinfection, never immerse the transducers longer than 45 minutes.
Damage may occur to the transducer housing and/or components if disinfection times exceed
these recommended limits.
Precautions – Panoramic (EFOV) Imaging
•
Measurements on a panoramic image are limited in accuracy by the acquisition of the image by
the user
•
Measurements out of plane are potentially less accurate.
•
A good B Mode/2D image is important for a high-quality panoramic image.
Precautions – 3D/4D Imaging
•
3D: Measurements out of plane are potentially less accurate due to movement of tissue.
•
A good B Mode/2D image is important for a high-quality 3D/4D image.
Precautions - WIRELESS
•
WIRELESS: The Mindray Wireless option device is an RF Receiver and Transmitter, operating
using industry standard 802.11 b/g protocols. Use of the wireless option, in the presence of
other high energy RF radiating devices may interfere with the transmission of data to the
network interfaces.
•
WIRELESS: The Mindray Wireless option device supports several industry standard
security protocols and should be enabled to reduce the chance of patient data, transmitted
over the wireless interface from the possible undesired interception of the data.
•
WIRELESS: Data transmission and reception rates are limited to by the bandwidth of the user’s
network infrastructure. A Quality of Service (QOS) level is determined by the number of users
on the network, the data being transferred by the users, distance of the ultrasound system to
the wireless access point and other factors.
Содержание Zonare ZS3
Страница 1: ...I ZS3 Ultrasound System Instructions for Use ...
Страница 6: ......
Страница 12: ......
Страница 38: ......
Страница 56: ...ZS3 Instructions for Use 48 P a g e Figure 4 6 P8 3m Adult and Pediatric TEE Scan Plane adjustment ...
Страница 74: ......
Страница 78: ......
Страница 86: ......
Страница 110: ......
Страница 128: ......
Страница 146: ......
Страница 150: ......
Страница 158: ......
Страница 160: ......
Страница 164: ...ZS3 Instructions for Use 147 P a g e ...
Страница 165: ...P N 046 012347 01 3 0 ...