ZS3 Instructions for Use
88
|
P a g e
iii. Avoid vessels, bile ducts, lung and ribs
iv. Avoid deep or long breath holds
5. Upon quiet inspiration and hold, wait 2-3 seconds for the image to stabilize
6. Single Frame 2D SWI is the default.
a. Press the Select key to update the 2D SWI ROI.
b. Press Select again to reacquire new 2D SWI data.
7. Once desired data has been collected, press the Freeze key.
8. Scroll back through images to select frames for calculation
9. Quality is displayed onscreen using two methods:
a. Quality map is available via OLED. Select the pane and toggle to the Quality map
using the Data Type OLED.
i. While Quality map is active, there is also a Reliability Index displayed
onscreen which takes into consideration, poor transducer contact, scan plane,
presence of artifact, rib, lung, bile ducts and blood vessels.
ii. The RLB Index can be removed from onscreen display by pressing RLB Index
OLED to toggle Off.
b. Motion Stability Index (M-STB Index) can be displayed onscreen
i. Based on a ‘five star’ quality rating with 5 stars representing exceptional
quality
ii. This index takes into consideration, motion from patient, probe, cardiac,
respiratory or anatomy/pathology movement.
iii. Can be removed from onscreen display by pressing M-STB Index OLED to
toggle Off.
10. Press Calc key and select desired calculation.
11. Press Store key to enter data into the Report
2D SWI Control
Description/Use
Dynamic
Single Frame
•
Dynamic 2D SWI automatically fires subsequent pulses to produce a series of 2D
SWI images.
•
Single Frame allows user to manipulate image and then manually press a key to
produce a single frame 2D SWI image.
Velocity/Stiffness
Quality
B (BMode Only)
Turn the OLED dial to select data type to be displayed on the active pane.
Data Type Swap
L to R
R to L
Toggle to switch data displayed from Right to Left Pane and Left to Right Pane.
Opacity
Turn the dial to adjust the degree of grayscale that is mixed with the 2D SWI overlay.
The range available is from 0-5:
•
0: least color overlay opacity
•
5: solid color overlay.
Unit
m/s, kPa, or m/s+kPa
Select unit of measure to be displayed on 2D SWI color bar and Calcs Results Boxes.
•
m/s: Shear wave velocity Cs
•
kPa: Elastic Module (Young’s Module)
M-STB Index
On/Off
Turn display of M-STB Index On and Off screen.
Содержание Zonare ZS3
Страница 1: ...I ZS3 Ultrasound System Instructions for Use ...
Страница 6: ......
Страница 12: ......
Страница 38: ......
Страница 56: ...ZS3 Instructions for Use 48 P a g e Figure 4 6 P8 3m Adult and Pediatric TEE Scan Plane adjustment ...
Страница 74: ......
Страница 78: ......
Страница 86: ......
Страница 110: ......
Страница 128: ......
Страница 146: ......
Страница 150: ......
Страница 158: ......
Страница 160: ......
Страница 164: ...ZS3 Instructions for Use 147 P a g e ...
Страница 165: ...P N 046 012347 01 3 0 ...