A-9
ST Low
-2.0 mV to (high limit - 0.2 mV)
-2.0 mV to 2.0 mV
A.7.2 Resp
Technique Trans-thoracic
impedance
Lead
Options are lead I and II. The default is lead II.
Respiration excitation
waveform
<300 μA RMS, 64 kHz (±10%)
Baseline impedance range
200 to 2500 (using an ECG cable with 1k resistance)
Bandwidth
0.2 to 2.5 Hz (-3 dB)
Sweep speed
3 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s, or 50 mm/s
Accuracy: ±5%
Respiration Rate
Measurement range
Adult:
Pediatric, neonate:
0 to 120 rpm
0 to 150 rpm
Resolution 1
rpm
Accuracy
7 to 150 rpm:
0 to 6 rpm:
±2 rpm or ±2%, whichever is greater
Not specified.
Apnea alarm time
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Alarm limit
Range (rpm)
Step (rpm)
RR High
Adult, pediatric:
Neonate:
(low limit + 2) to 100
(low limit + 2) to 150
1
RR Low
0 to (high limit – 2)
Содержание uMEC10
Страница 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 24: ...1 6 FOR YOUR NOTES ...
Страница 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Страница 36: ...2 12 FOR YOUR NOTES ...
Страница 50: ...3 14 FOR YOUR NOTES ...
Страница 58: ...4 8 FOR YOUR NOTES ...
Страница 72: ...6 8 FOR YOUR NOTES ...
Страница 118: ...8 28 FOR YOUR NOTES ...
Страница 174: ...18 8 FOR YOUR NOTES ...
Страница 186: ...20 6 FOR YOUR NOTES ...
Страница 192: ...21 6 FOR YOUR NOTES ...
Страница 202: ...23 4 FOR YOUR NOTES ...
Страница 206: ...24 4 FOR YOUR NOTES ...
Страница 224: ...26 10 FOR YOUR NOTES ...
Страница 248: ...B 8 FOR YOUR NOTES ...
Страница 266: ...C 18 FOR YOUR NOTES ...
Страница 280: ...E 6 FOR YOUR NOTES ...
Страница 286: ...F 6 FOR YOUR NOTES ...
Страница 287: ......
Страница 288: ...PN 046 008784 00 1 0 ...