22-3
22.4 Nurse Call
The patient monitor provides a nurse call connector to output nurse call signal when a
userdefined alarm occurs. To obtain nurse call signal, use the nurse call cable (
PN:
009-003116-00
) we supply to connect the hospital nurse call system to the nurse call
connector of the monitor and then follow this procedure:
1. Select
[
Main Menu
][
Maintenance >>
][
User Maintenance >>
]enter the
required password.
2. Select
[
Others >>
] to access the [
Others
] menu.
3. Select
[
Nurse Call Setup >>
] to change the nurse call settings as follows:
Select [
Signal Type
] and toggle between [
Pulse
] and [
Continuous
].
[
Pulse
]: the nurse call signals are pulse signals and each pulse lasts 1 second. When
multiple alarms occur simultaneously, only one pulse signal is outputted. If an
alarm occurs but the previous one is not cleared yet, a new pulse signal will also be
outputted.
[
Continuous
]: the nurse call signal lasts until the alarm ends, i.e. the duration of a
nurse call signal equals to that of the alarm condition.
Select [
Contact Type
] and toggle between [
Normally Open
] and [
Normally Closed
].
[
Normally Open
]: select if your hospital’s nurse call relay contact is normally
open.
[
Normally Closed
]: select if your hospital’s nurse call relay contact is normally
closed.
Select [
Alm Lev
] and set the alarm level for nurse call-triggering alarms.
Select [
Alarm Cat
.] and then select the category to which the nurse call-triggering
alarms belong.
Alarm conditions are indicated to nurses only when:
The nurse call system is enabled,
An alarm that meets your preset requirements occurs, and
The monitor is not in the alarm paused or reset status.
WARNING
z
To obtain the nurse call signal, use the nurse call cable (
PN: 009-003116-00
) we
supply. Otherwise the nurse call function will not work and the monitor may be
damaged.
z
Do not rely exclusively on the nurse call system for alarm notification. Remember
that the most reliable alarm notification combines audible and visual alarm
indications with the patient’s clinical condition.
Содержание uMEC10
Страница 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 24: ...1 6 FOR YOUR NOTES ...
Страница 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Страница 36: ...2 12 FOR YOUR NOTES ...
Страница 50: ...3 14 FOR YOUR NOTES ...
Страница 58: ...4 8 FOR YOUR NOTES ...
Страница 72: ...6 8 FOR YOUR NOTES ...
Страница 118: ...8 28 FOR YOUR NOTES ...
Страница 174: ...18 8 FOR YOUR NOTES ...
Страница 186: ...20 6 FOR YOUR NOTES ...
Страница 192: ...21 6 FOR YOUR NOTES ...
Страница 202: ...23 4 FOR YOUR NOTES ...
Страница 206: ...24 4 FOR YOUR NOTES ...
Страница 224: ...26 10 FOR YOUR NOTES ...
Страница 248: ...B 8 FOR YOUR NOTES ...
Страница 266: ...C 18 FOR YOUR NOTES ...
Страница 280: ...E 6 FOR YOUR NOTES ...
Страница 286: ...F 6 FOR YOUR NOTES ...
Страница 287: ......
Страница 288: ...PN 046 008784 00 1 0 ...