E.8 Patient Auxiliary Current
Patient Auxiliary currents are measured between any selected Applied Part connector and the
remaining Applied Part connector s. All measurements may have a true RMS only response.
The following outlet conditions apply when performing the Patient Auxiliary Current test.
normal polarity (Normal Condition)
reverse polarity (Normal Condition)
normal polarity with open neutral (Single Fault Condition)
reverse polarity with open neutral (Single Fault Condition)
normal polarity with open earth (Single Fault Condition)
reverse polarity with open earth (Single Fault Condition)
LIMITS
For CF
applied parts,
10 A in Normal Condition
50 A in Single Fault Condition
For BF
applied parts,
100 A in Normal Condition
500 A in Single Fault Condition
NOTE
z
Make sure the safety analyzer is authorized comply with requirement of
IEC60601-1.
z
Follow the instructions of the analyzer manufacturer.
Содержание uMEC10
Страница 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 24: ...1 6 FOR YOUR NOTES ...
Страница 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Страница 36: ...2 12 FOR YOUR NOTES ...
Страница 50: ...3 14 FOR YOUR NOTES ...
Страница 58: ...4 8 FOR YOUR NOTES ...
Страница 72: ...6 8 FOR YOUR NOTES ...
Страница 118: ...8 28 FOR YOUR NOTES ...
Страница 174: ...18 8 FOR YOUR NOTES ...
Страница 186: ...20 6 FOR YOUR NOTES ...
Страница 192: ...21 6 FOR YOUR NOTES ...
Страница 202: ...23 4 FOR YOUR NOTES ...
Страница 206: ...24 4 FOR YOUR NOTES ...
Страница 224: ...26 10 FOR YOUR NOTES ...
Страница 248: ...B 8 FOR YOUR NOTES ...
Страница 266: ...C 18 FOR YOUR NOTES ...
Страница 280: ...E 6 FOR YOUR NOTES ...
Страница 286: ...F 6 FOR YOUR NOTES ...
Страница 287: ......
Страница 288: ...PN 046 008784 00 1 0 ...