D-2
Measurement Alarm messages
L
Cause and solution
Extreme Brady
H
R on T
M*
Run PVCs
L*
PVCs/min
M*
Multif.PVC
M*
Bigeminy
M*
Trigeminy
M*
Tachy
M*
Brady
M*
Vent. Rhythm
M*
Multif. PVC
M*
Nonsus. Vtac
M*
Pause
L*
Resp
Resp Apnea
H
The respiration signal was so weak that the
monitor cannot perform respiration analysis.
Check the patient’s condition and the Resp
connections.
Resp Artifact
H
The patient’s heartbeat has interfered with
his respiration. Check the patient’s condition
and the Resp connections.
SpO
2
SpO
2
Desat
H
The SpO2 value has fallen below the
desaturation alarm limit. Check the patient’s
condition and check if the alarm limit
settings are correct.
No Pulse
H
The pulse signal was so weak that the
monitor cannot perform pulse analysis.
Check the patient’s condition, SpO
2
sensor
and measurement site.
CO
2
CO
2
Apnea
H
The patient stops breathing, or the
respiration signal was so weak that the
monitor cannot perform respiration analysis.
Check the patient’s condition and the RM
connections.
Содержание uMEC10
Страница 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 24: ...1 6 FOR YOUR NOTES ...
Страница 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Страница 36: ...2 12 FOR YOUR NOTES ...
Страница 50: ...3 14 FOR YOUR NOTES ...
Страница 58: ...4 8 FOR YOUR NOTES ...
Страница 72: ...6 8 FOR YOUR NOTES ...
Страница 118: ...8 28 FOR YOUR NOTES ...
Страница 174: ...18 8 FOR YOUR NOTES ...
Страница 186: ...20 6 FOR YOUR NOTES ...
Страница 192: ...21 6 FOR YOUR NOTES ...
Страница 202: ...23 4 FOR YOUR NOTES ...
Страница 206: ...24 4 FOR YOUR NOTES ...
Страница 224: ...26 10 FOR YOUR NOTES ...
Страница 248: ...B 8 FOR YOUR NOTES ...
Страница 266: ...C 18 FOR YOUR NOTES ...
Страница 280: ...E 6 FOR YOUR NOTES ...
Страница 286: ...F 6 FOR YOUR NOTES ...
Страница 287: ......
Страница 288: ...PN 046 008784 00 1 0 ...