A-3
Data storage
Operating mode
Continuous monitoring
Spot-checking
Capacity
96 hours of data
4000 data
Resolution
2 s
30 s
Stored data
Animal ID, animal category, SpO2 and PR value,
measurement time
Measurement specifications
SpO2
Range
0 to 100%
Resolution 1%
Accuracy
70 to 100%:
70 to 100%:
0% to 69%:
±2% (measured without motion)
±3% (measured with motion)
Unspecified
Refreshing rate
1 s
Averaging time
7 s (When the sensitivity is set to High)
9 s (When the sensitivity is set to Med)
11 s (When the sensitivity is set to Low)
PR
Range
18 to 300 bpm
Resolution 1
bpm
Accuracy
±
3 bpm (measured without motion)
±
5 bpm (measured with motion)
Refreshing rate
1 s
Averaging time
7 s (When the sensitivity is set to High)
9 s (When the sensitivity is set to Med)
11 s (When the sensitivity is set to Low)
Alarm limit specifications
Alarm limits
Range (% )
Step (% )
SpO
2
high limit
(low limit +1) to 100
SpO
2
low limit
50 to (high limit -1)
1
Alarm limits
Range (bpm)
Step (bpm)
PR high limit
(low limit +1) to 300
PR low limit
18 to (high limit -1)
1
Содержание PM-60Vet
Страница 1: ...PM 60Vet Pulse Oximeter Operator s Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 10: ...VIII FOR YOUR NOTES ...
Страница 12: ...2 FOR YOUR NOTES ...
Страница 30: ...2 8 FOR YOUR NOTES ...
Страница 34: ...3 4 FOR YOUR NOTES ...
Страница 56: ...6 6 FOR YOUR NOTES ...
Страница 64: ...7 8 FOR YOUR NOTES ...
Страница 78: ...10 2 FOR YOUR NOTES ...
Страница 82: ...A 4 FOR YOUR NOTES ...
Страница 96: ...P N 0852 20 77486 1 1 ...