6-3
6.3 Applying the Sensor
1. Select an appropriate SpO
2
sensor according to the animal category and weight.
2. Apply the SpO
2
sensor to the patient.
3. Connect the SpO
2
extension cable to the pulse oximeter.
6.4 Changing SpO
2
Settings
6.4.1 Switching On/Off SpO
2
and PR Alarms
1. Select
[
Menu
]
→
[
Alarm Setup
].
2. Select
SpO
2
or PR [
Alarm
] and toggle between:
[
On
]: When an alarm limit is exceeded, the pulse oximeter will give an alarm
according to the defined alarm level.
[
Off
]: When an alarm limit is exceeded, the pulse oximeter will not give an
alarm; the audible and visible alarm as well as the alarm message will be
switched off and the symbol
will be shown in the SpO
2
or PR area.
6.4.2 Setting Alarm Level
1. Select
[
Menu
]
→
[
System
]
→
[
Maintenance >>
]
→
enter required password.
2. Select
[
Alarm >>
].
3. Select
[
SpO2 Alm Lev
] and toggle between [
High
] and [
Med
].
4. Set
[
PR Alm Lev
] to [
High
], [
Med
] or [
Low
].
The level of the “No Pulse” alarm is configured to high and is not user adjustable.