4-6
4.10 Entering/Exiting the Standby Mode
4.10.1 Entering the Standby Mode
In the spot-checking mode, in situations where reviewing or exporting trend data is not
performed, the pulse oximeter will automatically enter the standby mode if no button
operation and SpO2 signal are detected for 1 minute and no "Battery Too Low" alarm
occurs. In the standby mode,
The backlight is turned off.
The standby screen is displayed.
Alarms excetp the “Battery too Low” alarm are disabled.
In the standby mode, if the “Battery too Low” alarm occurs, the system gives an audible
alarm which is defined as follows:
Double beep
Alarm volume: 2
Alarm interval: 30 seconds.
If you exit the standby mode at this time, the “Battery too Low” alarm will be given in
the form of techinical alarm.
4.10.2 Exiting the Standby Mode
In the standby mode, if the SpO
2
signal is detected or any button except the Power
button is pressed, the pulse oximeter will automatically exit the standby mode and
return to the previous mode.