4-2
4.2 Admitting a Patient
WARNING
z
Be sure to select correct patient category setting for your patient before
measurement. Wrong patient category may result in patient hazard due
to improper alarm limits.
4.2.1 Continuous Monitoring Mode
In the continuous monitoring mode, to admit a patient,
1. Select
[
Menu
]
→
[
Patient Info.
].
2. Set up [
Patient ID
] and [
Patient Cat.
].
3. Press the Left button to confirm the settings.
If you do not change the patient ID for a new patient, the patient will not be admitted.
4.2.2 Spot-checking Mode
In the spot-checking mode, to admit a patient,
1. Select
[
Menu
]
→
[
Patient Info.
].
2. Select
[
Patient Cat.
] and toggle between [
Adu
], [
Ped
] and [
Neo
].
3. Apply the SpO
2
sensor to the patient. After valid SpO
2
signals are detected,
The patient ID automatically changes to 1 if the current patient ID is 0. Thus
the patient is admitted.
The patient ID flashes and automatically increases by 1 after 8 seconds if the
current patient ID is not 0. Thus the patient is admitted.
If you press the Left button when the current patient ID is flashing, the patient ID will
stop flashing and remain unchanged. The patient will not be admitted and new
measurements will be stored under the current patient ID.
Содержание PM-60Vet
Страница 1: ...PM 60Vet Pulse Oximeter Operator s Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 10: ...VIII FOR YOUR NOTES ...
Страница 12: ...2 FOR YOUR NOTES ...
Страница 30: ...2 8 FOR YOUR NOTES ...
Страница 34: ...3 4 FOR YOUR NOTES ...
Страница 56: ...6 6 FOR YOUR NOTES ...
Страница 64: ...7 8 FOR YOUR NOTES ...
Страница 78: ...10 2 FOR YOUR NOTES ...
Страница 82: ...A 4 FOR YOUR NOTES ...
Страница 96: ...P N 0852 20 77486 1 1 ...