A-1
A
Product Specifications
Safety specifications (classified according to IEC60601-1)
Type of protection against
electrical shock
II (Internally powered equipment)
Degree of protection
against electric shock
Type BF – Applied part (defibrillation proof)
Degree of protection
against hazards of
explosion
Ordinary equipment, not protected
Degree of protection
against ingress of liquid
IPX2
Equipment type
Handheld
Mode of operation
Continuous
Physical specifications
Width × Height × Depth
56×124×30 mm
Max. weight
< 300g (full configuration, including the batteries)
Environmental
specifications
Operating conditions
Storage conditions
Temperature (ºC)
0 to 40
-20 to 60
Relative humidity
(non-condensing)
15% to 95%
10% to 95%
Atmospheric pressure
(mmHg)
425 to 809
120 to 809
Charger stand
Input voltage
100 to 240 VAC, 50/60 Hz
Output voltage
5 VDC
Output current
1.2 A
Output power
6 W