7-7
The structure of Byte 6 is:
7
6
5
4
3
2
1
0
Byte 6
0
0
For future use
LP
NS
SO
SPL
The value of LP is 1 or 0; 1 means low perfusion. The value of NS is 1 or 0; 1 means
that SpO
2
sensor is not connected. The value of SO is 1 or 0; 1 means SpO
2
sensor is off.
The value of SPL is 1 or 0; 1 means that the pulse searching time is too long.
7.3.2 Exporting Trend Data
You can export trend data through the multifunctional connector. To do so, follow this
procedure:
1. Connect a personal computer to the multifunctional connector using a
communication cable.
2. Select
[
Menu
]
→
[
Trend
] to enter the trend window.
3. Press the Left button to enter the [
Trend Setup
] menu.
4. Set
[
Export Port
] to [
Wire
].
5. Select
[
Export Trend
] to enter the trend exporting window.
When the trend exporting window is entered, all operations except exiting the trend
export window and turning off the pulse oximeter are disabled. To exit the trend
exporting window, press the Right button and select [
OK
].
Refer to the PC data management system software for detailed information on trend
data exporting.
Содержание PM-60Vet
Страница 1: ...PM 60Vet Pulse Oximeter Operator s Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 10: ...VIII FOR YOUR NOTES ...
Страница 12: ...2 FOR YOUR NOTES ...
Страница 30: ...2 8 FOR YOUR NOTES ...
Страница 34: ...3 4 FOR YOUR NOTES ...
Страница 56: ...6 6 FOR YOUR NOTES ...
Страница 64: ...7 8 FOR YOUR NOTES ...
Страница 78: ...10 2 FOR YOUR NOTES ...
Страница 82: ...A 4 FOR YOUR NOTES ...
Страница 96: ...P N 0852 20 77486 1 1 ...