T
ROUBLESHOOTING
- 64 -
PROBLEM
POSSIBLE CAUSES
WHAT TO DO
The system may have been
in STANDBY mode for a
while.
The Product resistivity is
below the Resistivity
Setpoint set in menu option
C02.
¾
Recirculate the water in the
system. See
Section 5-6
How to Recirculate Water
Manually before
Dispensing.
¾
Replace the SmartPak. See
Section 6-2 How to Replace
the SmartPak.
UV System only
The UV Timer is exhausted.
¾
Replace the UV Lamp. See
Section 6-9 How to Replace
the UV Lamp.
¾
After a new UV Lamp has
been installed, reset the UV
Timer. See
Section 6-10
How to View or Reset the
UV Lamp Timer (C05).
UV System only
The UV Lamp is not
installed correctly or not
installed at all.
¾
Power OFF the system and
reinstall the UV Lamp.
¾
If the red UV Lamp Alarm
is still displayed, then
contact Millipore.