U
SING THE
D
IRECT
-Q
- 31 -
F
ILLING
T
ANK
FILLING TANK mode is displayed when the tank
is being filled with RO water until the 100% Tank
Level display. FILLING TANK mode is launched
automatically when the Tank Level display is below
the 60% level or after a FLUSH cycle has been
completed.
FILLING TANK mode
Water can be dispensed or can be periodically
recirculated during FILLING TANK mode if the
Tank Level display is above the 10% level.
Dispensing or Recirculation during
FILLING TANK mode
P
RE
O
PERATE
PRE OPERATE mode is displayed when the system
is not dispensing water and not in FILLING TANK
mode. The Tank Level display is between the 60%
level and TANK FULL level.
The system will have a 3-minute auto-recirculation
every 2 hours in PRE OPERATE mode. The
Distribution Pump turns on. This will enhance the
quality of the Product Water delivered from the
system. Recirculation can also be activated manually
for up to 3 minutes.
The resistivity and temperature of the Product Water
is displayed during recirculation. The resistivity and
temperature remain displayed for up to 10 seconds
after recirculation is finished.
Содержание DIRECT-Q 3 SYSTEM
Страница 1: ...User Manual DIRECT Q 3 SYSTEM DIRECT Q 3 UV SYSTEM...
Страница 58: ...MAINTENANCE 50 G H...
Страница 60: ...MAINTENANCE 52 M...
Страница 63: ...MAINTENANCE 55 G H...