M
AINTENANCE
- 46 -
6-6
H
OW TO
E
MPTY THE
T
ANK
(C03)
The tank can be fully emptied of water through the POU Dispenser. This option is used when
performing a system sanitization. Before emptying the tank, it is recommended to remove the Final
Filter and to install the Barbed Fitting and Tygon Tubing. Place the other end of the Tygon Tubing in a
sink.
1.
Press the Main and
“+“ Buttons
together to enter
the menu. The
Display will show
C01.
2.
Press the Main
Button 2 times.
The Display will
show C03.
3.
Press the “+” and
“-“ Buttons
together. The
system will
dispense all the
water in the tank
through the POU
Dispenser. The
Tank Level display
will be blinking.
NOTE: To stop emptying the tank, press the Dispensing Button.