M
AINTENANCE
- 49 -
S
ANITIZING THE
S
YSTEM AND THE
T
ANK
Open the front cover to let the system go into STANDBY mode (C).
Remove the Vent Filter and install the clear elbow fitting from the Sanitization Kit (D).
Introduce the male connector of the clear tubing firmly into the elbow fitting (E).
Inject 200 ml (1 ml = 1 cc) of Hydrogen Peroxide solution (30%) into the 6 Litre tank via the clear
tubing (F).
Rinse the air vent port with 200 ml of purified water.
Close the front cover (G). The system will now go into FILLING TANK mode. Let the tank fill up to
the 100% level (H).
C
D
E
F
Содержание DIRECT-Q 3 SYSTEM
Страница 1: ...User Manual DIRECT Q 3 SYSTEM DIRECT Q 3 UV SYSTEM...
Страница 58: ...MAINTENANCE 50 G H...
Страница 60: ...MAINTENANCE 52 M...
Страница 63: ...MAINTENANCE 55 G H...