M
AINTENANCE
- 45 -
4.
Open the
Tank Outlet
Valve. Allow
the tank to be
emptied to its
lowest water
level.
NOTE: The lowest water level is not calibrated if the amount of water emptied from the tank
is less than 10%.
5.
Close the
Tank Outlet
Valve.
6.
Press the “+”
and “-“
Buttons
together to
enter the
lowest water
level. The
Display will
exit the menu
option and go
into
FILLING
TANK mode.
The Tank Level is now calibrated.
NOTE: If the Display remains in the menu option and the highest tank level is blinking, then
the tank level was not calibrated. The amount of water emptied from the tank was not enough
(less than 10%). The tank needs to be emptied to its lowest water level. Allow the tank to fill to
the 100% level and perform the calibration again.