38
: OPERation: STEP <numeric>
This command selects the test step: desired test step for single-step test and the
first step for multi-step test
Example: OPER: STEP 2
Example description: Select the test step STEP2
: OPERation: STEP? Return value: 2 <2
nd
STEP>
OPERation:FILE?
This command queries the current file name
Example:OPER:FILE? Return data: 26983877 <file nameis
26983877>
: RESUlt?
This command queries the test results
Example: RESU? Return value: 01,+9.00,+7.52562E-07,+1.07,2
with contents in the following sequence: test step ID, voltage measurement,
current measurement, resistance measurement, determination results
where determination results are predefined codes with the meanings defined in the
table below:
Code Description
1
Test aborted
2
Determination PASS
3
Measurement value over upper limit
(determination FAIL)
4
Measurement value below lower limit
(determination FAIL)
5
Arc detection (determination FAIL)
6
Breaking point voltage (determination FAIL)
7
Voltage error (the difference between the
measured voltage and the voltage set is too
great)
8
Open circuit error
: STARt
This command starts a test which functions the same as the TEST key in the front
panel
Example: STARt.
Содержание HT-7630
Страница 1: ...HI POT Tester 7630 User s Manual Version 1 2 Issued Date March 2017...
Страница 33: ...29 6 4 Connection cable and method...
Страница 44: ...40 panel Example TEST ABOR...