26
7
FAIL
O
Determination FAIL
8
HV-ON
O
High voltage output in progress
9
GND
--- Chassis Ground
10
TRIG
I
Test trigger
11
ABORT
I
Test abortion
12
RESET
I
Reset your machine
13
+5V
--- DC +5V output
14
+12V
--- DC +12Voutput
15
COM
--- Circuit common terminal
Use description:
PIN1-PIN8 is a collector open output without any voltage or signal. Current passing
through the connector must be less than 50mA.
(1) TRIG (Pin10): Functions the same as the TEST key in the front panel when short
circuited with Pin15 (COM)
(2) ABORT (Pin11): Functions the same as the ABORT key in the front panel when
short circuited with Pin15 (COM)
(3) PASS (Pin6): Conduct with Pin15 (COM) when the DUT is determined to PASS by
your machine
(4) FAIL (Pin7): Conduct with Pin15 (COM) when the DUT is determined to FAIL by
your machine
(5) TESTING (Pin5): Conduct with Pin15 (COM) when your machine is testing
(6) H.V. ON: Conduct with Pin15 (COM) when your machine is outputting high
voltage