judgment over the arc during the test.
(10)
OFFS:
This item resets the test end to prevent the impact of the external jig on
accuracy. It deducts the leak current by the jig in advance and displays the actual
leak current of the DUT with the equation “measurement readings = actual
measurement less reset value”. This item has two options: Manual and auto. The
first requires manual input of the leak current value. The second contains the
following steps: Remove the DUT from your machine; point the cursor to the field
OFFS and press the software key AUTO; your machine measures according to the
given voltage statistics and prompts the results; the message “SAVE=ENTER”
displays after the measurement operation is done; press the ENTER key to save
the test value or the EXIT key to exit without saving it; the saved value will be
deducted from each measurement to enable a more accurate test.