6
2.3Rear Panel
(1)
High voltage output terminal:
Connection to the expansion box
(2)
Low voltage output terminal:
Connection to the expansion box
(3)
USB port:
Connection to the computer for controlling the machine with
computer instructions
(4)
Expansion box port:
Connect this port and an expansion box with a connection
cable
(5)
AC power socket:
AC power input connector with a 250V/3.15A fuse
(6)
Iron casing terminal:
Ground your machine with this terminal for more steady
device operation
(7)
RS-232 connector:
Connect to the computer to control your machine with
computer instructions
(8)
High voltage tester port:
Connect the high voltage tester to this port
(9)
Remote controller port:
Signal for external control including the determination
and testing signal output or input of an external signal for test start and stop
(10)
Inter Lock port:
High voltage current output safety switch. Short circuit both
ends of this port during testing for the test voltage output. Opening the circuit of
this port during testing will stop the test and prompt the message “INTER LOCK”
(11)
Exhaust outlet:
Air exhaust for cooling
(12)
Serial number sticker:
Indicates the serial number of your machine
(1)
(2)
(3) (4)
(5)
(6)
(7)
(9)
(10
(11)
(12)
(8)