28
6
RS-232/USB device
6.1
RS- 232 interface
Baud Rate: Available options are 9600/19200/38400/57600/115200 and it can
be set up in the System mode
Transmission bits: 1 initial bit, 8 data bits, 1 ending bit
6.2
Command format
The RS-232 interface function of your machine is exercised by a command string
in ASCII code for remote control and setup. Each command string is composed of
the command code and its parameters, two commands are connected with a
semicolon “;”, and appended with an end code. There are two types of end code:
LF(0x0a) and CR(0x0d)+LF(0x0a).
6.3
Connector
Your machine employs a 9-pin male connector with pin functions outlined in the
table below:
Pin
Name Description
1
CD
Reserved
2
RxD
Receive data
3
TxD
Transmit data
4
DTR
Reserved
5
GND
Ground point
6
DSR
Reserved
7
RTD
Reserved
8
CTS
Reserved
9
RI
Reserved