Page 88
Data Processing
thermoIMAGER TIM Connect
6.7
Relative Extreme Values
Within the
Tools
,
Configuration
and
Extended measuring
menu it is possible to exclude pixel within the
thermal image from data analysis.
Using
Ignore pixels with temperatures
a temperature can be set as lowest (
below
) or highest
(
above
) value.
In case of analyzing an image using measure areas (
for measure areas
) the calculation of e.g. the mean
value is not taking values below or above the set limits into account.
For hot/cold spots localization
determines that all pixels with the defined values are excluded from
search.
For </> temperature ranging
means that all pixels with the respective temperature do not influence
e.g. the automatic scaling of temperature-time-diagram.
Fig. 87 Relative Extreme values