Software Configuration
thermoIMAGER TIM Connect
On the analog inputs you can select each function only once (exception: Uncommitted value).
On the analog outputs also each function can only be selected once (exception: Measurement area; Alarm).
i
You can use the analog inputs, outputs and the digital input of the PIF simultaneously.
Fig. 42 View configuration
If you activate the function Alarm on an analog output you need to assign also in the Alarms tab the desired
measurement area to the desired PIF output.
The industrial PIF has a maximum of three analog outputs. To use additional outputs, the ADAM-4024 module
from Advantech is recommended, which has four analog outputs per module and can be cascaded with
further modules.
For more information on the settings for alarms, see Chap.
6.8
.