Setting Sensor Parameters
optoNCDT 1900 / Profinet
7.6.2
Behavior of the Distance Values
Master values based on the zeroing or master function are coded with 18 bits. The master can assume twice the measuring range.
The examples demonstrate the behavior of the digital value with an ILD1900-100-IE, measuring range 100 mm.
Target with 16% of the measuring range
Target with 60% of the measuring range
Target with 60% of the measuring range
SMR
EMR
0 %
16 %
100 %
MR
Target
108718
16.00 mm
SMR
EMR
0 %
60 %
100 %
MR
Target
137554
60.00 mm
SMR
EMR
0 %
60 %
100 %
MR
Target
137554
60.00 mm
Zero setting (master value = 0 mm)
Zero setting (master value = 0 mm)
Setting master value 10 mm
SMR
EMR
0 %
16 %
100 %
MR
Target
98232
0.00 mm
87746
-16.00 mm
153282
84.00 mm
EMR
SMR
0 %
60 %
100 %
MR
Target
98232
0.00 mm
58910
-60.0 mm
124446
40.00 mm
EMR
SMR
0 %
60 %
100 %
MR
Target
104786
10.00 mm
65464
-50.0 mm
131000
50.00 mm