MDM300 I.S. User’s Manual
48
97213 Issue 5, October 2017
APPENDIX A
Appendix A
Technical Specifi cations
Performance
Measurement Technology
Michell Ceramic Sensor
Accuracy
±1°C from -60 to +20°Cdp (-76 to +68°Fdp)
±2°C from -100 to -60°Cdp (-148 to -76°Fdp)
±0.2°C (0.3°F) temperature
Calibrated Measurement Range
For Spot Checks -70 to +20°Cdp (-95 to +68°Fdp)
For Online Analysis -100 to -70°Cdp (-148 to -95°Fdp)
Uncalibrated Readings From
+20 to +30°Cdp (+68 to +86°Fdp)
Measurement Units
°C, °F, K dew point & gas temperature
% RH, ppm
V
, g/m
3
ppm
W
& g/kg (for air, N
2
, H
2
, CO
2
, SF
6
, user-de
fi
ned mw)
ppm
V
, lb/mmscf & g/m
3
for natural gas
Pressure Units
psig, barg, KPa, bara, psia, MPa
Resolution (display)
0.1 for all dew-point derived units & auto ranging where appropriate
Resolution (measurement)
Better than 0.1°Cdp (0.2°Fdp)
Typical Response Speed
T95 in
≤
30 minutes to -60°Cdp (-76°Fdp)
Sample Gas Flow
0.2 to 0.5 Nl/min (0.5 to 1 scfh)
Electrical Specifi cations
Auxiliary Inputs
Connection to Michell Easidew I.S. or Easidew PRO I.S. via Remote
Sensor Interface
Battery
4.8 V internally mounted Nickel Metal Hydride (NiMH) battery
Battery Operating Life
Up to 24 hours of typical usage between charges
Battery Charger
Intelligent charger, NOT certi
fi
ed for hazardous area use
Charger supply voltage: 100 to 240 V AC, 50/60 Hz, 8 VA
Operating Conditions
Operating Pressure
0 to 35 MPa (0 to 350 barg / 0 to 5076 psig)
Operating Environment
Outdoors 0 to +100% RH (non-condensing)
Operating Temperature
-20 to +50°C (-4 to +122°F)
Storage / Transport Temperature
-40 to +70°C (-40 to +158°F)
Altitude
Up to 2000m (6562ft)
Mechanical Specifi cations
Display
Blue LCD graphical display
Enclosure Type
Steel
fi
ber-loaded high-impact polyamide 6
IP/NEMA Rating
IP66/NEMA 4
Gas Connections
1/8” NPT female (other options available)
Gas Wetted Materials
AISI 316L stainless steel, PTFE Seal, Borosilicate glass, ceramic
Outline Dimensions
218 x 170 x 90mm (8.6 x 6.7 x 3.5ft) (d x w x h)
Weight
1.5kg (3.3lbs)
Содержание MDM300 I.S
Страница 1: ...MDM300 I S Advanced Dew Point Hygrometer User s Manual I S 97213 Issue 5 October 2017...
Страница 55: ...MDM300 I S User s Manual Michell Instruments 47 APPENDIX A Appendix A Technical Specifications...
Страница 58: ...MDM300 I S User s Manual 50 97213 Issue 5 October 2017 APPENDIX B Appendix B Datalog Status Display...
Страница 61: ...MDM300 I S User s Manual Michell Instruments 53 APPENDIX C Appendix C Hazardous Area Certification...
Страница 64: ...MDM300 I S User s Manual 56 97213 Issue 5 October 2017 APPENDIX D Appendix D FCC Declaration...
Страница 70: ...MDM300 I S User s Manual 62 97213 Issue 5 October 2017 NOTES...
Страница 71: ......
Страница 72: ...http www michell com...