MDM300 I.S. User’s Manual
Michell Instruments
33
OPERATION
3.6
Guide to Measurement and Sampling
Operating the MDM300 I.S. is very straightforward, but sampling conditions can vary
greatly from one application to another. To get the best possible response speed, the
appropriate procedure should be followed.
!
HIGH PRESSURE! High pressure gases are potentially
hazardous. Energy stored in these gases can be released
suddenly and with extreme force. High pressure systems
should be assembled and operated only by people who have
been trained in proper safety practices.
!
POSSIBLE INJURY! The tubing, valves and other apparatus
attached to this instrument must be adequate for the
maximum pressure which will be applied, otherwise physical
injury to the operator or bystander is possible.
For the most basic measurement it is recommended to have a flow meter (0 to 1 l/min),
and at least one metering valve.
A number of sampling kits are available from Michell Instruments:
•
The Easi-fit sample kit is a low cost sampling system consisting
of 2 metering valves and a flow meter which is suitable for many
applications.
•
The MDM300 panel-mount sampling system offers a complete package
for conditioning of a sample, prior to measurement with an MDM300.
It is contained within an optional flight case which allows for easy
transport of everything required to make the measurements. The anti-
static construction of the case makes it suitable for use in hazardous
areas.
Figure 27
Easi-Fit Sample Kit MDM300 Panel-Mount Sampling System
Specialized kits are available for the measurement of atmospheric pressure samples,
compressed air, medical gas and SF
6
. Contact a Michell Instruments’ representative for
more information.
Содержание MDM300 I.S
Страница 1: ...MDM300 I S Advanced Dew Point Hygrometer User s Manual I S 97213 Issue 5 October 2017...
Страница 55: ...MDM300 I S User s Manual Michell Instruments 47 APPENDIX A Appendix A Technical Specifications...
Страница 58: ...MDM300 I S User s Manual 50 97213 Issue 5 October 2017 APPENDIX B Appendix B Datalog Status Display...
Страница 61: ...MDM300 I S User s Manual Michell Instruments 53 APPENDIX C Appendix C Hazardous Area Certification...
Страница 64: ...MDM300 I S User s Manual 56 97213 Issue 5 October 2017 APPENDIX D Appendix D FCC Declaration...
Страница 70: ...MDM300 I S User s Manual 62 97213 Issue 5 October 2017 NOTES...
Страница 71: ......
Страница 72: ...http www michell com...