MDM300 I.S. User’s Manual
Michell Instruments
45
CALIBRATION
6.3
Calibration Correction Method
NOTE: Any logging procedure currently in operation must be cancelled before
the calibration procedures can be entered.
To enter the CALIBRATION Menu page, highlight
CALIBRATION
from the SET-UP Menu
by using the
or
keys and press the
Enter
key.
Entry to the CALIBRATION Menu is protected by a passcode (
4876
).
Once the passcode has been successfully entered, the following page will be shown.
POINT
CALIBRATION
-100
-90
-80
-70
-60
REF
-100.0
-90.0
-80.0
-70.0
-60.0
MDM300
-100.0
-90.0
-80.0
-70.0
-60.0
Figure 31
Calibration Menu Page
The MDM300 I.S. Calibration look-up table is divided into three columns.
The values in the first column
POINT
represent the thirteen nominal calibration points,
spread at 10°C intervals over the instruments operating range of -100 to +20°C (-148
to +68°F
)
. These points cannot be edited.
The second column
REF
represents the value of the reference dew-point standard
applied at that particular point.
The third column
MDM300
represents the reading that the MDM300 I.S. was showing
at the corresponding
REF
level input.
The values in both the
REF
and
MDM300
I.S. columns can be edited by means of the
and
keys. The
and
keys provide the means of switching between corresponding
table columns.
After changes have been made press the
Enter
key to reach the Exit page. No changes
are implemented until the
key is pressed at the Exit page.
To exit without implementing any changes, press the
Enter
key to reach the Exit page.
Press the
key, from the Exit page, to return to the SET-UP Menu.
Содержание MDM300 I.S
Страница 1: ...MDM300 I S Advanced Dew Point Hygrometer User s Manual I S 97213 Issue 5 October 2017...
Страница 55: ...MDM300 I S User s Manual Michell Instruments 47 APPENDIX A Appendix A Technical Specifications...
Страница 58: ...MDM300 I S User s Manual 50 97213 Issue 5 October 2017 APPENDIX B Appendix B Datalog Status Display...
Страница 61: ...MDM300 I S User s Manual Michell Instruments 53 APPENDIX C Appendix C Hazardous Area Certification...
Страница 64: ...MDM300 I S User s Manual 56 97213 Issue 5 October 2017 APPENDIX D Appendix D FCC Declaration...
Страница 70: ...MDM300 I S User s Manual 62 97213 Issue 5 October 2017 NOTES...
Страница 71: ......
Страница 72: ...http www michell com...