MDM300 I.S. User’s Manual
Michell Instruments
39
GOOD MEASUREMENT PRACTICE
4.1 Sampling
Hints
Measurement of moisture content is a complex subject, but does not need to be difficult.
This section aims to explain the common mistakes made in measurement situations, the
causes of the problem, and how to avoid them. Mistakes and bad practices can cause
the measurement to vary from the expectation; therefore a good sampling technique is
crucial for accurate and reliable results.
Transpiration and Sampling Materials
- 20
- 30
- 40
- 50
- 60
- 70
1
2
3
4
5
Time
(
hours
)
Dew point
(
ºC
)
nylon
copper
polyethylene
PTFE
nickel
stainless steel
All materials are permeable to water vapor, as the water molecule is extremely small
compared to the structure of solids, even when compared to the crystalline structure of
metals. The graph to the right shows the dew point inside tubing of different materials
when purged with very dry gas, where the exterior of the tubing is in the ambient
environment.
Many materials contain moisture as part of their structure, particularly organic materials
(natural or synthetic), salts (or anything which contains them) and anything which
has small pores. It is important to ensure that the materials used are suitable for the
application.
If the partial water vapor pressure exerted on the outside of a compressed air line is
higher than on the inside, the atmospheric water vapor will naturally push through the
porous medium causing water to migrate into the pressurized air line. This effect is
called transpiration.
Adsorption and Desorption
Adsorption is the adhesion of atoms, ions, or molecules from a gas, liquid, or dissolved
solid to the surface of a material, creating a film. The rate of adsorption is increased at
higher pressures and lower temperatures.
Desorption is the release of a substance from or through the surface of a material. In
constant environmental conditions, an adsorbed substance will remain on a surface
almost indefinitely. However, as the temperature rises, so does the likelihood of
desorption occurring.
In practical terms, as the temperature of the environment fluctuates, water molecules
are adsorbed and desorbed from the internal surfaces of the sample tubing, causing
small fluctuations in the measured dew point.
Содержание MDM300 I.S
Страница 1: ...MDM300 I S Advanced Dew Point Hygrometer User s Manual I S 97213 Issue 5 October 2017...
Страница 55: ...MDM300 I S User s Manual Michell Instruments 47 APPENDIX A Appendix A Technical Specifications...
Страница 58: ...MDM300 I S User s Manual 50 97213 Issue 5 October 2017 APPENDIX B Appendix B Datalog Status Display...
Страница 61: ...MDM300 I S User s Manual Michell Instruments 53 APPENDIX C Appendix C Hazardous Area Certification...
Страница 64: ...MDM300 I S User s Manual 56 97213 Issue 5 October 2017 APPENDIX D Appendix D FCC Declaration...
Страница 70: ...MDM300 I S User s Manual 62 97213 Issue 5 October 2017 NOTES...
Страница 71: ......
Страница 72: ...http www michell com...