3.4
SET-UP Menu Parameters
Follow the instructions in Section 3.3.1 to select the required parameters.
3.4.1
SETTINGS
PRIMARY DP AT
ATMOS
DP/TEMP UNIT
°C
PRESSURE UNIT
Psig
GAS TYPE
AIR
MOL WEIGHT
N/A
CHART INTERVAL
5 s
SETTINGS
Figure 15
SETTINGS Page
Parameter
Description
PRIMARY DP AT
Indicates if the primary display sensor is at atmospheric or system
pressure.
Available Input:
ATMOS
or
PRESS
(Atmospheric or Pressure)
DP/TEMP UNIT
Sets system temperature units.
Available Input:
°C, K, °F
PRESSURE UNIT
Sets system pressure units.
Available Input:
psig, MPa, psia, bara, KPa, barg,
GAS TYPE
De
fi
nes the sample gas type.
NOTE
: The
MOL WEIGHT
fi
eld automatically reads
N/A
for all
gases other than
USER.
For any other type of sample gas, the molecular weight of the gas
must be de
fi
ned in the
MOL WEIGHT
fi
eld.
Available Input:
Air, USER, H2, SF6, CO2, N2
MOL WEIGHT
Sets molecular weight of user-de
fi
ned sample gas, default value 1.
Available Input: Range
1 to 99
CHART INTERVAL
Sets chart time interval, default value 5.
Available Input: Range
1 to 60 seconds
Table 4
SETTINGS Parameters
Содержание MDM300 I.S
Страница 1: ...MDM300 I S Advanced Dew Point Hygrometer User s Manual I S 97213 Issue 5 October 2017...
Страница 55: ...MDM300 I S User s Manual Michell Instruments 47 APPENDIX A Appendix A Technical Specifications...
Страница 58: ...MDM300 I S User s Manual 50 97213 Issue 5 October 2017 APPENDIX B Appendix B Datalog Status Display...
Страница 61: ...MDM300 I S User s Manual Michell Instruments 53 APPENDIX C Appendix C Hazardous Area Certification...
Страница 64: ...MDM300 I S User s Manual 56 97213 Issue 5 October 2017 APPENDIX D Appendix D FCC Declaration...
Страница 70: ...MDM300 I S User s Manual 62 97213 Issue 5 October 2017 NOTES...
Страница 71: ......
Страница 72: ...http www michell com...