![Michell Instruments MDM300 I.S Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/michell-instruments/mdm300-i-s/mdm300-i-s_user-manual_1784666034.webp)
MDM300 I.S. User’s Manual
26
97213 Issue 5, October 2017
OPERATION
3.4.4
EXTERNAL (Sensor Interface)
!
The MDM300 I.S. is designed to be connected ONLY to a
Michell Instruments’ remotely mounted Easidew I.S. or
Easidew PRO I.S. transmitter. Connection to any other type of
sensor could impair safety.
The Easidew I.S. or Easidew PRO I.S. must be connected
retaining the associated I.S. barriers and using a Michell
Remote Sensor Interface module. See Section 2.6.
EXTERNAL TYPE
NONE
EXTERNAL ZERO
N/A
EXTERNAL SPAN
N/A
USER PRESSURE
ATM
Psig
EXTERNAL
Figure 19
EXTERNAL SET-UP Page
Parameter
Description
EXTERNAL TYPE
De
fi
nes use of Easidew I.S. transmitter.
NOTE: The ZERO and SPAN limits are automatically added
for this transmitter.
Available Input:
NONE,
EASIDEW
EXTERNAL ZERO
De
fi
nes external zero (lower limit).
The
ZERO
default limit is pre-programmed but can be rede
fi
ned.
Available Input: Range:
-120 to +30°C
EXTERNAL SPAN
De
fi
nes external span (upper limit).
The
SPAN
default limit is pre-programmed but can be rede
fi
ned.
Available Input: Range:
-120 to +30°C
USER PRESSURE
Used in the calculation of dew point at pressure, moisture content
and absolute humidity. If a user pressure is not selected then the
MDM300 I.S. assumes atmospheric pressure. (Refer to Section 2.6.2.)
NOTE: Pressure compensation does not affect the reading
of the external sensor.
Available Input: Range
ATM
,
0 to 300 barg
(or equivalent in other
units).
Table 7
EXTERNAL Sensor Parameters
Содержание MDM300 I.S
Страница 1: ...MDM300 I S Advanced Dew Point Hygrometer User s Manual I S 97213 Issue 5 October 2017...
Страница 55: ...MDM300 I S User s Manual Michell Instruments 47 APPENDIX A Appendix A Technical Specifications...
Страница 58: ...MDM300 I S User s Manual 50 97213 Issue 5 October 2017 APPENDIX B Appendix B Datalog Status Display...
Страница 61: ...MDM300 I S User s Manual Michell Instruments 53 APPENDIX C Appendix C Hazardous Area Certification...
Страница 64: ...MDM300 I S User s Manual 56 97213 Issue 5 October 2017 APPENDIX D Appendix D FCC Declaration...
Страница 70: ...MDM300 I S User s Manual 62 97213 Issue 5 October 2017 NOTES...
Страница 71: ......
Страница 72: ...http www michell com...