Liquidew EExd User’s Manual
76
97097 Issue 14.1, November 2019
APPENDIX H
H.4
Special Conditions of Use
1. Do not open when an explosive gas atmosphere may be present.
2. External cables shall be compatible with a temperature of 93°C (T5) or 109°C (T4).
3. Maximum process pressure shall not exceed 80 barg. Maximum permissible
process pressure is marked on the equipment.
4. Maximum permitted sample flow rate into the enclosure shall not exceed
0.3LPH.
5. All process lines shall be purged to ensure the process gas or liquid is above
its upper explosive limit before applying power.
6. Where painted or powder coated, the enclosures could present an electrostatic
hazard. Clean only with a damp or anti-static cloth.
7. The enclosure is to be earthed externally using the earth point provided.
8. Only suitably ATEX / IECEx certified (as appropriate) cable glands and blanking
elements shall be used.
Refer to the relevant sections within this manual for the connection, wiring and cable
glanding requirements.
H.5
Maintenance and Installation
The Liquidew EExd must only be installed by suitably qualified personnel and in
accordance with the instructions provided and the terms of the applicable product
certificates.
Maintenance and servicing of the product must only be carried out by suitably trained
personnel or returned to an approved Michell Instruments’ Service Center.
Flame paths are not intended to be repaired.
Содержание Liquidew EExd
Страница 1: ...Liquidew EExd Process Moisture Analyzer User s Manual 97097 Issue 14 1 November 2019...
Страница 42: ...Liquidew EExd User s Manual 34 97097 Issue 14 1 November 2019 APPENDIX A Appendix A Technical Specifications...
Страница 48: ...Liquidew EExd User s Manual 40 97097 Issue 14 1 November 2019 APPENDIX B Appendix B Variable Definitions...
Страница 53: ...Liquidew EExd User s Manual Michell Instruments 45 APPENDIX C Appendix C Software...
Страница 55: ...Liquidew EExd User s Manual Michell Instruments 47 APPENDIX D Appendix D Calibration Correction...
Страница 60: ...Liquidew EExd User s Manual 52 97097 Issue 14 1 November 2019 APPENDIX E Appendix E Modbus RTU Communications...
Страница 70: ...Liquidew EExd User s Manual 62 97097 Issue 14 1 November 2019 APPENDIX F Appendix F Modbus RTU Details...
Страница 77: ...Liquidew EExd User s Manual Michell Instruments 69 APPENDIX G Appendix G Register Number Formats...
Страница 85: ...Liquidew EExd User s Manual Michell Instruments 77 APPENDIX I Appendix I Quality Recycling Warranty Information...
Страница 89: ...Liquidew EExd User s Manual Michell Instruments 81 NOTES...
Страница 90: ...http www michell com...