Liquidew EExd User’s Manual
Michell Instruments
13
INSTALLATION
1 2 3 4 5 6
NC C NO NC C NO
7 8 9 10 11 12
13 14 15
16 17 18 19
20 21 22 23
NC C NO NC C NO
AL3
AL2
AL1
AL4
COMMS
Channel 1
Channel 2
Optional
PL16
PL20
PL15
PL14
PL19
B A GND
mA1+
mA1-
mA2+
mA2-
mA3+
mA3-
mA4+
mA4-
L E N
Gland
Cable screen
Enclosure wall
{
Figure 6
Hook-up Wiring Diagram
2.5.3 Analog and Digital Communications
Two active 4-20 mA outputs and a Modbus RS485 digital interface (see Appendix F for
details) are provided with the Liquidew EExd. mA1 and mA3 output the moisture or
dew-point values. mA2 and mA4 output the sample temperature values of Channels 1
& 2 respectively.
NOTE: The maximum output resistance for the 4-20 mA outputs is 500Ω.
See Section 3.7 on the setting of the 4-20 mA outputs via the user interface and
Appendix F on setting the outputs via the Modbus interface.
Refer to
Figure 7 for cable wiring.
2.5.4 Process Alarms and Analyzer Status Alarms
Each channel has an associated process and fault alarm, as shown below:
AL1: Channel 1 process alarm
AL2: Channel 1 analyzer status alarm
AL3: Channel 2 process alarm (Optional)
AL4: Channel 2 analyzer status alarm (Optional)
The process alarm contacts can be set as either Normally Closed (
N/C
) or Normally
Open (
N/O
).
The process alarm contacts change from
N/C
to
N/O
when the moisture value becomes
greater than the alarm set point.
The fault alarm contacts change from
N/C
to
N/O
when an error associated with the
channel occurs or when there is a supply failure.
Содержание Liquidew EExd
Страница 1: ...Liquidew EExd Process Moisture Analyzer User s Manual 97097 Issue 14 1 November 2019...
Страница 42: ...Liquidew EExd User s Manual 34 97097 Issue 14 1 November 2019 APPENDIX A Appendix A Technical Specifications...
Страница 48: ...Liquidew EExd User s Manual 40 97097 Issue 14 1 November 2019 APPENDIX B Appendix B Variable Definitions...
Страница 53: ...Liquidew EExd User s Manual Michell Instruments 45 APPENDIX C Appendix C Software...
Страница 55: ...Liquidew EExd User s Manual Michell Instruments 47 APPENDIX D Appendix D Calibration Correction...
Страница 60: ...Liquidew EExd User s Manual 52 97097 Issue 14 1 November 2019 APPENDIX E Appendix E Modbus RTU Communications...
Страница 70: ...Liquidew EExd User s Manual 62 97097 Issue 14 1 November 2019 APPENDIX F Appendix F Modbus RTU Details...
Страница 77: ...Liquidew EExd User s Manual Michell Instruments 69 APPENDIX G Appendix G Register Number Formats...
Страница 85: ...Liquidew EExd User s Manual Michell Instruments 77 APPENDIX I Appendix I Quality Recycling Warranty Information...
Страница 89: ...Liquidew EExd User s Manual Michell Instruments 81 NOTES...
Страница 90: ...http www michell com...