Liquidew EExd User’s Manual
50
97097 Issue 14.1, November 2019
APPENDIX D
D.3
Calibration Procedure
!
Caution: A valid correction of the calibration can only
be done by a recognized humidity calibration laboratory
accredited by a standards institution such as UKAS or NIST.
Any calibration performed by an unrecognized body will
invalidate the calibration of the instrument.
1. Press the
Get Table
button to download the correction table for either
Channel 1 or 2 and check to see if the values are equal and exact. If
not, then the instrument has had its calibration adjusted previously. To
simplify the correction procedure, reset the adjusted values back to their
‘equal & exact’ default values.
2. Connect the Liquidew EExd to a suitable dew-point generator with a
reference hygrometer (preferably chilled-mirror type with traceable
calibration). Purge the Liquidew for a minimum of 5 days with air that is
drier than -100°Cdp (-148°Fdp) before commencing the calibration.
3. Increase the generator set-point in steps of 10°Cdp (18°Fdp), moving
sequentially up the Liquidew’s measurement range from –100 to +20°Cdp
(-148 to +68°Fdp) and record the dew point measured by the reference
and the Liquidew.
Suitable stabilization periods should be allowed between changing the
generator setting and taking the final sensor count and reference hygrometer
values. Actual stabilization times will depend to a degree on the type of
generator used, the gas flow rate and the integrity of the pipe work and
sensor manifold. As a rule of thumb, it is strongly suggested that the following
stabilization times are taken as a minimum requirement:
Dew-point Temperature
°Cdp
Minimum Stabilization Time
-100
5 days
-90
12 hours
-80
10 hours
-70
8 hours
-60
4 hours
-50
2 hours
-40
1 hour
-30
1 hour
-20
1 hour
-10
1 hour
0
1 hour
+10
1 hour
+20
1 hour
If calibration is not being done to -100°Cdp (-148°Fdp) then the purge times
can be reduced:
Drier than -70°Cdp (-94°Fdp)
5 days
-41 to -69°Cdp (-41.8 to -92°Fdp)
64 hours
+20 to -40°Cdp (+68 to -40°Fdp)
16 hours
Содержание Liquidew EExd
Страница 1: ...Liquidew EExd Process Moisture Analyzer User s Manual 97097 Issue 14 1 November 2019...
Страница 42: ...Liquidew EExd User s Manual 34 97097 Issue 14 1 November 2019 APPENDIX A Appendix A Technical Specifications...
Страница 48: ...Liquidew EExd User s Manual 40 97097 Issue 14 1 November 2019 APPENDIX B Appendix B Variable Definitions...
Страница 53: ...Liquidew EExd User s Manual Michell Instruments 45 APPENDIX C Appendix C Software...
Страница 55: ...Liquidew EExd User s Manual Michell Instruments 47 APPENDIX D Appendix D Calibration Correction...
Страница 60: ...Liquidew EExd User s Manual 52 97097 Issue 14 1 November 2019 APPENDIX E Appendix E Modbus RTU Communications...
Страница 70: ...Liquidew EExd User s Manual 62 97097 Issue 14 1 November 2019 APPENDIX F Appendix F Modbus RTU Details...
Страница 77: ...Liquidew EExd User s Manual Michell Instruments 69 APPENDIX G Appendix G Register Number Formats...
Страница 85: ...Liquidew EExd User s Manual Michell Instruments 77 APPENDIX I Appendix I Quality Recycling Warranty Information...
Страница 89: ...Liquidew EExd User s Manual Michell Instruments 81 NOTES...
Страница 90: ...http www michell com...