Liquidew EExd User’s Manual
54
97097 Issue 14.1, November 2019
APPENDIX E
E.4
Termination Resistor
A 120R termination resistor is provided on the instrument’s circuit board for systems
that require it. To connect the resistor, place the link provided across JMP2.
E.5
Register Map
The following tables describe the instrument registers with their address location,
Modbus function and the number format.
Address
dec
Address
hex
Function
Read/
Write
Default
Value
Register
Config-
uration
0
0000*
Instrument Address
R/W
F
1
0001
Sensor 1 Humidity Value – Hi Word (dp/moisture)
R
L
2
0002
Sensor 1 Humidity Value – Lo Word (dp/moisture)
R
L
3
0003
Sensor 1 Ambient Temperature
R
H
4
0004
Status
R
D
5
0005
Sensor 2 Humidity Value – Hi Word (dp/moisture)
R
L
6
0006
Sensor 2 Humidity Value – Lo Word (dp/moisture)
R
L
7
0007
Time to next log Mins + Secs
R
I
8
0008
Sensor 2 Ambient Temperature
R
H
9
0009
Sensor 1 Flow Reading l/min
R
F
10
000A
Sensor 2 Flow Reading l/min
R
F
11
000B
Sensor 1 Humidity Setting (dp or moisture)
R/W
B
12
000C*
mA1 output maximum value (sensor 1 humidity)
R/W
K
13
000D*
mA1 output minimum value (sensor 1 humidity)
R/W
K
14
000E*
mA2 output maximum value (sensor 1
temp)
R/W
M
15
000F*
mA2 output minimum value (sensor 1 temp)
R/W
M
16
0010*
Sensor 2 Humidity Setting (dp or moisture)
R/W
B
17
0011*
mA3 output maximum value (sensor 2 humidity)
R/W
F
18
0012*
mA3 output minimum value (sensor 2 humidity)
R/W
K
19
0013*
Logging Interval
R/W
I
20
0014*
mA4 output maximum value (sensor 2 temp)
R/W
M
21
0015*
mA4 output minimum value (sensor 2 temp)
R/W
M
22
0016*
Sensor 1 Alarm Set Point
R/W
K
23
0017*
Sensor 2 Alarm Set Point
R/W
K
24
0018*
RTC Year (val1) + Month (val2)
R/W
I
25
0019*
RTC Date (val1) + Hours (val2)
R/W
I
26
001A*
RTC Mins (val1) + Secs (val2)
R/W
I
27
001B*
Sensor 1 Solute Ref Number
R/W
H
28
001C*
Sensor 2 Solute Ref Number
R/W
H
29
001D
Units/ Command
R/W
E
30
001E
Water DP Sensor 1 – Batch Number
R
I
31
001F
Water DP Sensor 1 – Serial Number
R
I
32
0020
Water DP Sensor 1 - Year
R
I
33
0021
Water DP Sensor 1 – Month and day
R
I
34
0022
Water DP Sensor 1– Hours Of Operation
R
F
35
0023
Error Indicator
R
C
36
0024
Not Used
Содержание Liquidew EExd
Страница 1: ...Liquidew EExd Process Moisture Analyzer User s Manual 97097 Issue 14 1 November 2019...
Страница 42: ...Liquidew EExd User s Manual 34 97097 Issue 14 1 November 2019 APPENDIX A Appendix A Technical Specifications...
Страница 48: ...Liquidew EExd User s Manual 40 97097 Issue 14 1 November 2019 APPENDIX B Appendix B Variable Definitions...
Страница 53: ...Liquidew EExd User s Manual Michell Instruments 45 APPENDIX C Appendix C Software...
Страница 55: ...Liquidew EExd User s Manual Michell Instruments 47 APPENDIX D Appendix D Calibration Correction...
Страница 60: ...Liquidew EExd User s Manual 52 97097 Issue 14 1 November 2019 APPENDIX E Appendix E Modbus RTU Communications...
Страница 70: ...Liquidew EExd User s Manual 62 97097 Issue 14 1 November 2019 APPENDIX F Appendix F Modbus RTU Details...
Страница 77: ...Liquidew EExd User s Manual Michell Instruments 69 APPENDIX G Appendix G Register Number Formats...
Страница 85: ...Liquidew EExd User s Manual Michell Instruments 77 APPENDIX I Appendix I Quality Recycling Warranty Information...
Страница 89: ...Liquidew EExd User s Manual Michell Instruments 81 NOTES...
Страница 90: ...http www michell com...