Liquidew EExd User’s Manual
Michell Instruments
25
OPERATION
3.7.5 User-Defined Solutes
When selecting the sample solute type it is also possible to select up to four different
user-defined solutes, labelled by default as U1, U2, U3 and U4. These are available by
scrolling through to the end of the list of sample solute types. The table of saturation
constants (or Cs values) of each user-defined solute can be edited by the user. From
the Variables page, to edit your solute, scroll down the solute list and choose one of the
user-defined solutes. Select the solute and it will show the the Cs value editor page.
Each user-defined solute may also be given a customized label which will be appended
to the end of the default labels.
3.7.5.1 Editing the Table of Cs Values
Variable
Brief Description
0°C
Cs Value at temp of 0
°C
10°C
Cs Value at temp of 10
°C
20°C
Cs Value at temp of 20
°C
30°C
Cs Value at temp of 30
°C
40°C
Cs Value at temp of 40
°C
50°C
Cs Value at temp of 50°C
To view the table of Cs values, scroll to the desired user-defined solute using the
Up
(
) and
Down (
) buttons, press the
SELECT
button and the relevant table will
appear.
To edit any of the values use the
Up
(
) and
Down (
) buttons to scroll up and down
the list. Scroll to the desired temperature and press the
SELECT
button. A small box
will appear beside the Cs value to indicate that it can be adjusted. Use the
Up
(
)
and
Down (
) buttons to change the value - these can be changed at a faster rate by
extending the duration of the arrow button press. Press the
SELECT
button again when
finished and to adjust another value.
Press the
MENU/MAIN
button to display the label editing page.
3.7.5.2 Editing the Label for the User-defined Solute
The custom label can be up to 8 characters in length and consists of any combination
of letters from A-Z, space, hyphen and numbers from 0 to 9.
To enter the page, press the
MENU/MAIN
button from the page showing the table of
Cs values. Use the
Up
(
) and
Down (
) buttons to select the character and press the
SELECT
button to save and move to the next character position.
To exit back to Variables Page 1, press the
MENU/MAIN
button.
Содержание Liquidew EExd
Страница 1: ...Liquidew EExd Process Moisture Analyzer User s Manual 97097 Issue 14 1 November 2019...
Страница 42: ...Liquidew EExd User s Manual 34 97097 Issue 14 1 November 2019 APPENDIX A Appendix A Technical Specifications...
Страница 48: ...Liquidew EExd User s Manual 40 97097 Issue 14 1 November 2019 APPENDIX B Appendix B Variable Definitions...
Страница 53: ...Liquidew EExd User s Manual Michell Instruments 45 APPENDIX C Appendix C Software...
Страница 55: ...Liquidew EExd User s Manual Michell Instruments 47 APPENDIX D Appendix D Calibration Correction...
Страница 60: ...Liquidew EExd User s Manual 52 97097 Issue 14 1 November 2019 APPENDIX E Appendix E Modbus RTU Communications...
Страница 70: ...Liquidew EExd User s Manual 62 97097 Issue 14 1 November 2019 APPENDIX F Appendix F Modbus RTU Details...
Страница 77: ...Liquidew EExd User s Manual Michell Instruments 69 APPENDIX G Appendix G Register Number Formats...
Страница 85: ...Liquidew EExd User s Manual Michell Instruments 77 APPENDIX I Appendix I Quality Recycling Warranty Information...
Страница 89: ...Liquidew EExd User s Manual Michell Instruments 81 NOTES...
Страница 90: ...http www michell com...