Liquidew EExd User’s Manual
Michell Instruments
65
APPENDIX F
The table below is an example of a message sent by the master, to read the time to
the next log (register 7) and Channel 2 sample temperature (register 8). The message
shows a master addressing a Liquidew EExd with a slave address of 01H and a Modbus
function of 03H that informs the Liquidew EExd that it wishes to read two registers
starting from address 07H. Bytes 3 & 4 hold the starting address and bytes 5 & 6 hold
the number of registers to be read. Bytes 7 & 8 contain the CRC code that is calculated
using bytes 1 to 6 as represented below.
Byte No
Meaning
Value
1
Slave Address
01H
2
Modbus function code
03H
3
Starting Address MSB
00H
4
Starting Address LSB
07H
5
No of points MSB
00H
6
No of points LSB
02H
7
CRC Lo Byte
??H
8
CRC Hi Byte
??H
Read Request Message
In response to the above message, the Liquidew EExd may transmit the following
message:
Byte No
Meaning
Value
1
Slave Address
01H
2
Modbus function code
03H
3
Byte Count
04H
4
Data MSB
05H
5
Data LSB
26H
6
Data MSB
00H
7
Data LSB
FAH
8
CRC Lo Byte
??H
9
CRC Hi Byte
??H
Read Response Message
This response repeats the address of the Liquidew EExd and the function code, along
with the byte count, the data and the CRC. In this example, the request asked for the
values from two registers. Therefore, the number of bytes returned is four. The value of
register 7 is contained in bytes 4 & 5 and the value of register 8 in bytes 6 & 7. Register
7 = 5m 26s and register 7 = 00FA = 252 = 25.2°C in this example. Any errors within
the data of Read Request Message will result in an exception being raised.
Содержание Liquidew EExd
Страница 1: ...Liquidew EExd Process Moisture Analyzer User s Manual 97097 Issue 14 1 November 2019...
Страница 42: ...Liquidew EExd User s Manual 34 97097 Issue 14 1 November 2019 APPENDIX A Appendix A Technical Specifications...
Страница 48: ...Liquidew EExd User s Manual 40 97097 Issue 14 1 November 2019 APPENDIX B Appendix B Variable Definitions...
Страница 53: ...Liquidew EExd User s Manual Michell Instruments 45 APPENDIX C Appendix C Software...
Страница 55: ...Liquidew EExd User s Manual Michell Instruments 47 APPENDIX D Appendix D Calibration Correction...
Страница 60: ...Liquidew EExd User s Manual 52 97097 Issue 14 1 November 2019 APPENDIX E Appendix E Modbus RTU Communications...
Страница 70: ...Liquidew EExd User s Manual 62 97097 Issue 14 1 November 2019 APPENDIX F Appendix F Modbus RTU Details...
Страница 77: ...Liquidew EExd User s Manual Michell Instruments 69 APPENDIX G Appendix G Register Number Formats...
Страница 85: ...Liquidew EExd User s Manual Michell Instruments 77 APPENDIX I Appendix I Quality Recycling Warranty Information...
Страница 89: ...Liquidew EExd User s Manual Michell Instruments 81 NOTES...
Страница 90: ...http www michell com...