4.3
Replacement of the Flow Meter
DANGER
Electric
Shock Risk
The power to the enclosure must be turned off before any
work is carried out in the measurement system enclosure.
Observe de-energize durations
3)
4)
1
2
Figure 25
Flow Meter Replacement
1. Isolate the incoming sample line by
CLOSING
the sample inlet isolation
valve and allow the system to depressurize.
NOTE: Always refer to
Appendix H.4 - Special Conditions of Safe Use.
2. Remove the enclosure window and mount the user interface (see Section
4.1).
3. Restrain the flow meter assembly (1) and use a 14mm wrench to loosen
and remove the 2 off 6mm sample pipe connections.
4. Unscrew the 2-wire cable connector (2) from the flow meter.
5. Carefully withdraw the flow meter assembly out of the enclosure.
6. Replacement of the flow meter is simply a reversal of the above procedure.
7. To ensure the full requirement of this product's safety certificate is
maintained, any loosened or disturbed tubework or couplings must be
subject to a gas pressure test and appropriate leak check at 1.5x the max
operating pressure before the full product is re-energized.
Содержание Liquidew EExd
Страница 1: ...Liquidew EExd Process Moisture Analyzer User s Manual 97097 Issue 14 1 November 2019...
Страница 42: ...Liquidew EExd User s Manual 34 97097 Issue 14 1 November 2019 APPENDIX A Appendix A Technical Specifications...
Страница 48: ...Liquidew EExd User s Manual 40 97097 Issue 14 1 November 2019 APPENDIX B Appendix B Variable Definitions...
Страница 53: ...Liquidew EExd User s Manual Michell Instruments 45 APPENDIX C Appendix C Software...
Страница 55: ...Liquidew EExd User s Manual Michell Instruments 47 APPENDIX D Appendix D Calibration Correction...
Страница 60: ...Liquidew EExd User s Manual 52 97097 Issue 14 1 November 2019 APPENDIX E Appendix E Modbus RTU Communications...
Страница 70: ...Liquidew EExd User s Manual 62 97097 Issue 14 1 November 2019 APPENDIX F Appendix F Modbus RTU Details...
Страница 77: ...Liquidew EExd User s Manual Michell Instruments 69 APPENDIX G Appendix G Register Number Formats...
Страница 85: ...Liquidew EExd User s Manual Michell Instruments 77 APPENDIX I Appendix I Quality Recycling Warranty Information...
Страница 89: ...Liquidew EExd User s Manual Michell Instruments 81 NOTES...
Страница 90: ...http www michell com...