930 Compact IC Flex SeS (2.930.1400)
6 Troubleshooting
Problem
Cause
Remedy
Marked drop in
pressure.
Leak in the system.
Check the capillary connections and seal leaks,
if necessary
.
The baseline has a
large amount of
noise.
The eluent path has a leak.
Check the eluent path and fix the leak.
If necessary, tighten the pressure screws using
a wrench (6.2739.000).
High-pressure pump – Con-
taminated pump valves.
Clean the pump valves
The eluent path is blocked.
Check the eluent path and eliminate the block-
age.
Contaminations in the elu-
ent
Check the quality of the chemicals and the
water.
High-pressure pump –
Defective piston seals.
MCS – The
CO
2
adsorber cartridge CW
is depleted.
Replace the CO
2
.
The pulsation absorber is
not connected or is defec-
tive.
Connect
or replace
the pulsation absorber.
MCS – The vacuum pump is
defective.
Contact Metrohm Service.
The baseline is drift-
ing.
Leak in the system.
Check all capillary connections and seal leaks,
if necessary
.
The organic solvent in the
eluent is evaporating.
■
■
Constantly stir the eluent.
The pressure in the
system markedly
increases.
The inline filter
(6.2821.120) is blocked.
Replace the filter (6.2821.130) .