5 Operation and maintenance
930 Compact IC Flex SeS (2.930.1400)
■■■■■■■■
63
3 Reinserting the outlet valve into the pump head
■
Insert the outlet valve into the outlet valve holder (the seal must
be visible) (
1
).
■
Screw the outlet valve holder up into the pump head (
2
) and
tighten it firmly by hand and then retighten it one additional
¼
turn using the adjustable wrench (
3
).
■
Tighten the connection capillary to the auxiliary piston back onto
the outlet valve holder.