Exchanging the sample loop
1
Pressure screw
Fastened to Port 6.
2
Sample loop
3
Pressure screw
Fastened to Port 3.
Exchanging the sample loop
NOTE
Observe a dead-volume-free installation of the sample loop
ing dead-volume-free capillary connections", page 15)
1 Removing the existing sample loop
■
Loosen the pressure screws (6.2744.010) at Port 3 and Port 6.
■
Remove the sample loop.
2 Installing a new sample loop
■
Fasten one end of the sample loop to Port 3 using a PEEK pressure
screw (6.2744.010).
■
Use the second PEEK pressure screw (6.2744.010) to fasten the
other end of the sample loop to Port 6.