3.18 Connecting the separation column
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
930 Compact IC Flex SeS (2.930.1400)
3 Connecting the guard column
CAUTION
When inserting the guard column, ensure that it is inserted cor-
rectly based on the marked flow direction (if specified).
■
Fasten the inlet of the guard column to the column inlet capillary
using a short pressure screw (6.2744.070).
■
If the guard column is connected to the separation column using
a connection capillary, fasten this connection capillary to the
guard column outlet with a pressure screw.
Rinsing the guard column
1 Rinsing the guard column
■
Place a beaker under the guard column's outlet.
■
Start manual control in MagIC Net and select the high-pressure
pump:
Manual
▶
Manual control
▶
Pump
–
Flow
:
in accordance with column leaflet
–
On
■
Rinse the guard column with eluent for approx. 5 minutes.
■
Stop the high-pressure pump in the manual control in MagIC Net
again:
Off
.
3.18
Connecting the separation column
The smart separation column (iColumn) is the heart of ion chromato-
graphic analysis. It separates the different components according to their
interactions with the column. Metrohm separation columns are equipped
with a chip where their technical specifications and history (start-up, oper-
ating hours, injections etc) are stored.
NOTE
Information regarding which separation column is suitable for your
application can be found in the
Metrohm Column Program
, the
product information for the separation column or it can be obtained
through your representative.
You can find product information for your separation column at
in the Ion Chromatography product area.