3 Installation
930 Compact IC Flex SeS (2.930.1400)
■■■■■■■■
35
Connecting the rinsing solution inlet with STREAM
1
Connect the detector outlet capillary of the conductivity detector and
the capillary labeled
rinsing solution
to each other using a coupling
(6.2744.040) and two pressure screws (6.2744.070).
NOTE
The detector outlet capillary must not be shortened.
Connecting the rinsing solution inlet to the Dosino
The
rinsing solution
capillary can be connected to the Dosino to which the
regenerant
capillary is already connected.
For this step, you need the following accessories: (not contained in its
entirety in the standard accessories):
■
3 short pressure screws (6.2744.070)
■
Inline filter (6.2821.120)
■
PTFE capillary (6.1803.030)
■
Adapter M6/UNF (6.2744.080)
■
FEP tubing M6 / 100 cm (6.1805.120)
■
Adapter Dosino Port 4, M6 inner (6.1808.280)