5 Operation and maintenance
930 Compact IC Flex SeS (2.930.1400)
■■■■■■■■
85
7 Connecting and conditioning the Metrohm Suppressor Mod-
ule (MSM)
■
Reconnect all capillaries of the MSM to the IC system.
■
Before switching the MSM over for the first time, rinse the three
suppressor units with solution for five minutes.
5.11
Servicing the Metrohm CO
2
Suppressor (MCS)
5.11.1
Replacing the CO
2
adsorption cartridge CW
Capacity loss
The adsorption capacity of the CO
2
adsorption cartridge CW is limited and
decreases over time. This is indicated by an increasing baseline (since more
CO
2
gets to the detector). The CO
2
adsorption cartridge CW has to be
replaced once the adsorption capacity is depleted. CO
2
adsorption car-
tridges CW are available under the order number 6.2837.100.
Maintenance interval
The CO
2
adsorption cartridge CW (6.2837.100) loses capacity over time.
Depleted cartridges change their color to violet. New cartridges are white.
The CO
2
adsorption cartridge CW needs to be replaced:
■
once
¾
of its color has changed to violet.
■
after 12 months at the latest.
Removing the depleted CO
2
adsorption cartridge CW
1
Remove the depleted CO
2
adsorption cartridge CW from the adsorp-
tion cartridge holder.
2
Remove the
Air in
capillary of the Metrohm CO
2
Suppressor (MCS).
3
Professionally dispose of the depleted CO
2
adsorption cartridge CW.
Installing the new CO
2
adsorption cartridge CW
1
Preparing the new CO
2
CO2 adsorption cartridge CW", page 38)
.
2
Installing the new CO
2
CO2 adsorption cartridge CW", page 39)
.