3 Installation
930 Compact IC Flex SeS (2.930.1400)
■■■■■■■■
41
4 Putting the adsorption cartridge holder into the instrument
■
Place the adsorption cartridge holder with the CO
2
adsorption car-
tridge CW into the instrument's detector chamber.
3.12
Installing the conductivity detector
The 930 Compact IC Flex provides enough space for one detector and
additional accessories in the detector chamber. The detector is available as
separate device and is supplied with a separate manual.
Placing the detector in the instrument
Follow the instructions in the chapter
Inserting the detector
in the manual
for the detector.
Connecting the detector to the eluent path
NOTE
The separation column is not inserted into the instrument until it is
being started up for the first time. Until then, the detector inlet capillary
has to be connected to the
out
outlet of the MCS using a long pressure
screw (6.2744.090).
Accessories
For this step, you need the following accessories:
■
Pressure screw, long (6.2744.090)