Page 6 of 20
A
B
C
Setting Up the Instrument
Unpacking the Instrument
1. Remove the instrument components from all packaging without damaging
the packing materials.
2. Save the original packaging materials to be used if the instrument must be shipped
for maintenance at a later date.
DiluCult™ Instrument
The shipping box contains the following:
• One dispensing assembly
(6.4 mm diameter for GL45 bottle per pump)
• One fixed arm assembly
• One set of 10 GeckoGrip
• One set of adhesive gum
• One monitoring software CD
• One power cord
• One USB cable
• One Allen wrench
• One quick-start guide
• One Certificate of Conformity
DiluCult™ 2 Instrument
The shipping box contains the following:
• One dispensing assembly
(6.4 mm diameter for GL45 bottle per pump)
• One set of 10 GeckoGrip
• One set of adhesive gum
• One monitoring software CD
• One power cord
• One USB cable
• One quick-start guide
• One Certificate of Conformity
• Optional accessories
Installing the Instrument
Place the instrument on a flat, stable, horizontal surface, away from any source of excessive
heat and close to an easily accessible, properly grounded power supply outlet.
Note: A flat, stable, horizontal, vibration-free surface is necessary for the accuracy
of the weighing platform.
Installing the Fixed Arm (DiluCult™ Instrument Only)
Install the fixed arm at the back of the instrument. Use the Allen wrench and two screws
provided. Choose between three possible positions: A, B, or C.