Page 4 of 20
Specifications
Category
DiluCult™ Instrument
DiluCult™ 2 Instrument
Cat. No.
5622900001
5623350001
Weighing range
0.1 g to 3,000 g
0.1 g to 5,000 g
Weighing accuracy
0 to 100 g: ± 0.05 g
100 g to 3,000 g: ± 0.1%
0 to 100 g: ± 0.05 g
100 g to 5,000 g: ± 0.1%
Minimum weight for dilution
5 g
5 g
Minimum weight for distribution
5 g
5 g
Minimal weight of the sample
1 g
1 g
Maximum authorized weight
3,000 g
5,000 g
Ideal functioning temperature for optimal
accuracy of the weighing platform
-10 °C / + 40 °C
-10 °C / + 40 °C
Accuracy of the instrument from 1 to 100 g
± 0.05 g
± 0.05 g
Accuracy of the instrument from 100 to 5,000 g
± 0.1%
± 0.1%
Maximun weight without support
4,500 g
7,000 g
Dilution factor
1
/
2
to
1
/
99
1/2 to 1/99
Dilution accuracy
> 98%
> 99%*
Dilution time for a 25 g sample diluted at 1/10
< 9 s**
< 9 s**
Dilution time for a 375 g sample diluted at 1/10
−
< 80 s**
Dispensing time for 90 mL
< 7 s**
< 7 s**
Dispensing time for 225 mL
< 10 s**
< 10 s**
Maximum number of pumps
2 integrated
2 integrated
Dispensing/distribution mode
Fast/Standard/Accurate
Fast/Standard/Accurate
Traceability
Printer/Monitoring software (data export to Microsoft
®
Excel
®
, OpenOffice
®
Calc,
LibreOffice
®
Calc, printer)
Exported data
Sample weight/Total weight/Dilution factor/Dilution accuracy/Right or left
pump + with monitoring software: Sample number/Name of operator/Date/Time
Connection
RS232 output: To connect a printer; USB output: To connect a computer;
Jack output: To connect a foot pedal
Weight
12 kg
14.2 kg
Dimensions (width x depth x height)
37 x 44 x 38 cm
14.6 x 17.3 x 14.9 in.
37 x 48 x 38 cm
14.6 x 18.9 x 14.9 in.
Power source
100–240 V ~± 10%
50 / 60 Hz ± 2 Hz 60 w
Fuse
T3.15A L 250V 5 x 20 mm
Installation category
II
IP
2
In compliance with
ISO 7218 / ISO 6887-1 / FDA-BAM
EI 61010-1: 2010
*In standard or accurate mode
**With booster kit in fast mode