20
Chapter 5: MEI Software
memory. The values stored in data memory are lost when the DSPpro is reset (F9 key) or the
power is turned off.
The reset function (F9) loads the firmware and configuration parameters from boot memory into
data memory. Values stored in boot memory will be read by the SETUP program on
initialization.
Selecting "
Save Defaults to File
" saves the current boot memory configuration to a disk file with
the extension .ABS. Selecting "
Load Defaults From Disk
" will load the values from a disk file
into the boot memory on board.
Screen Values
Save Defaults
Save Defaults to File
Load Defaults from File
Data Memory
Boot Memory
Disk File
Read Defaults
Screen Values are Read from
Data Memory
DefaultsTransferred from
Boot Memory to Data Memory
Data Transferred from
Data Memory to Boot Memory
Data Transferred from
Boot Memory to Disk
Data Transferred from
Disk File to Boot Memory
Figure 5-1. SETUP Parameter Storage